• Home
  • FAQ
  • More
    • Home
    • FAQ
  • Home
  • FAQ

Frequently Asked Questions

Why perform beam quality testing for x-ray imaging equipment used in radiology?

Why perform beam quality testing for x-ray imaging equipment used in radiology?

Why perform beam quality testing for x-ray imaging equipment used in radiology?

The beam quality test for x-rays (or gamma rays) is an important medical imaging equipment performance test. Defined by the half-value layer (HVL) and the x-ray tube's peak voltage potential used, it validates that there is enough beam filtration to remove low-energy photons.  Low-energy photons degrade  image quality which may then trigg

The beam quality test for x-rays (or gamma rays) is an important medical imaging equipment performance test. Defined by the half-value layer (HVL) and the x-ray tube's peak voltage potential used, it validates that there is enough beam filtration to remove low-energy photons.  Low-energy photons degrade  image quality which may then trigger the need for image retakes,  and ultimately in unnecessary additional radiation dose to the patient.  When monitored in a routine basis, beam quality testing may reveal problems with equipment's wear and tear such as problems with the x-ray tube or the high-voltage generator. The HVL is also an important parameter in radiation dosimetry calculations, dose conversion tables, etc. 

Why should I get the device?

Why perform beam quality testing for x-ray imaging equipment used in radiology?

Why perform beam quality testing for x-ray imaging equipment used in radiology?

The MuCT device enables you to:

  • perform HVL measurements for computed tomography (CT) equipment; on-demand/anytime
  • determine the x-ray beam's effective energy
  • perform measurements in true clinical conditions ; without the need to use the CT system service mode
  • establish a more complete/robust clinical medical physics equipment performance pro

The MuCT device enables you to:

  • perform HVL measurements for computed tomography (CT) equipment; on-demand/anytime
  • determine the x-ray beam's effective energy
  • perform measurements in true clinical conditions ; without the need to use the CT system service mode
  • establish a more complete/robust clinical medical physics equipment performance program
  • monitor HVL and effective energy changes as a function of time

What is the device made of?

Why perform beam quality testing for x-ray imaging equipment used in radiology?

Are there any other advantages to using the device?

The MuCT device is made of Aluminum 1100 alloy.

Are there any other advantages to using the device?

Are there any other advantages to using the device?

Are there any other advantages to using the device?

The MuCT device may be useful for:

  • generating required radiation dosimetry data such as the HVL, effective energy, and off-center bow-tie filter attenuation analysis
  • providing data inputs for computer-simulated radiation dosimetry calculations


Once properly researched and/or validated, the device may open the possibility to be used for:

  • PET/C

The MuCT device may be useful for:

  • generating required radiation dosimetry data such as the HVL, effective energy, and off-center bow-tie filter attenuation analysis
  • providing data inputs for computer-simulated radiation dosimetry calculations


Once properly researched and/or validated, the device may open the possibility to be used for:

  • PET/CT attenuation corrections
  • Dual-Energy CT HVL
  • Cone-Beam CT HVL

What is included in the introductory offer?

Are there any other advantages to using the device?

Do I need any especial equipment to use the device?

Our introductory offer includes:

  • The MuCT device
  • Supporting base
  • Carrying/storage case
  • Data analysis/calculations tool

Do I need any especial equipment to use the device?

Are there any other advantages to using the device?

Do I need any especial equipment to use the device?

The MuCT device is designed to accommodate the pencil-type ionization chambers currently used in medical physics equipment performance testing.  The device's inner diameter is designed to match the diameter of the holes in the CTDI phantom. If your equipment includes a sleeve to add to the chamber when performing CTDI measurements, then you should also use the sleeve with the MuCT device. 

The x-ray tubes used in CT already have heavy inherent filtration. Is this device useful at all?

Computer simulations/modeling such as Monte Carlo can be used to estimate the HVL, effective energy, and radiation dose estimates, etc. Does this device has any value?

Computer simulations/modeling such as Monte Carlo can be used to estimate the HVL, effective energy, and radiation dose estimates, etc. Does this device has any value?

You are correct, x-ray tubes used in CT already have heavy inherent filtration. The MuCT device is useful in allowing for a more complete and inclusive clinical medical physics performance testing program. It may also allow for the early detection of problems with the x-ray tube and/or the high-voltage generator, or other system  componen

You are correct, x-ray tubes used in CT already have heavy inherent filtration. The MuCT device is useful in allowing for a more complete and inclusive clinical medical physics performance testing program. It may also allow for the early detection of problems with the x-ray tube and/or the high-voltage generator, or other system  components. Lastly, there's the benefit of using the device to obtain input parameters for computer models/simulations used in CT dosimetry calculations.

Computer simulations/modeling such as Monte Carlo can be used to estimate the HVL, effective energy, and radiation dose estimates, etc. Does this device has any value?

Computer simulations/modeling such as Monte Carlo can be used to estimate the HVL, effective energy, and radiation dose estimates, etc. Does this device has any value?

Computer simulations/modeling such as Monte Carlo can be used to estimate the HVL, effective energy, and radiation dose estimates, etc. Does this device has any value?

Computer simulations are widely used in CT radiation dosimetry to estimate patient dose. However, simulation results still need to be validated. The MuCT device provides the capabilities to validate computer simulation input/output data or possibly improve it.  Moreover, the MuCT device enables physicists to obtain empirical measurements that are unique to each CT system of interest.

How does the device compares to Monte Carlo computer calculations?

Computer simulations/modeling such as Monte Carlo can be used to estimate the HVL, effective energy, and radiation dose estimates, etc. Does this device has any value?

How does the device compares to Monte Carlo computer calculations?

The MuCT device:

  • has no need for manufacture's proprietary/confidential information
  • allows for empirical measurements to calculate HVL and effective energy
  • can be used for measurements specific to different CT system manufacturers, models, and different systems of the same model/type
  • is light/portable
  • is easy to use
  • does not requires computer s

The MuCT device:

  • has no need for manufacture's proprietary/confidential information
  • allows for empirical measurements to calculate HVL and effective energy
  • can be used for measurements specific to different CT system manufacturers, models, and different systems of the same model/type
  • is light/portable
  • is easy to use
  • does not requires computer simulation/model programing skills

Whereas Monte Carlo:

  • need manufacture's proprietary/confidential design data
  • generates simulated HVL and effective energy results
  • output simulations are limited to one CT system vendor and one model, capture differences of systems of the same model
  • is not portable
  • is not easy to program
  • requires computer simulation/model programing skills

What evidence do you have that his device works as presented?

Isn't this device outdone by the "one-shot" HVL measurements using solid state detectors?

How does the device compares to Monte Carlo computer calculations?

During our research we found that the broad-beam geometry HVL as measured with the device, and using its corresponding effective energy, resulted in an experimental narrow-beam geometry HVL that is within 5% percent difference from the manufacture's narrow-beam geometry HVL.

What about Compton scatter and buildup?

Isn't this device outdone by the "one-shot" HVL measurements using solid state detectors?

Isn't this device outdone by the "one-shot" HVL measurements using solid state detectors?

Our research results strongly support the conclusion that the MuCT device's design, combined with the symmetrical x-ray irradiation, creates the conditions to achieve ideal broad-beam geometry and ideal broad-beam attenuation as described by Evans in "The Atomic Nucleus" and Attix in "Introduction to Radiological Physics and Radiation Dos

Our research results strongly support the conclusion that the MuCT device's design, combined with the symmetrical x-ray irradiation, creates the conditions to achieve ideal broad-beam geometry and ideal broad-beam attenuation as described by Evans in "The Atomic Nucleus" and Attix in "Introduction to Radiological Physics and Radiation Dosimetry".  Furthermore, we don't see the typical Compton scatter buildup "shoulder" when we plot the attenuation as a function of aluminum thickness.

Isn't this device outdone by the "one-shot" HVL measurements using solid state detectors?

Isn't this device outdone by the "one-shot" HVL measurements using solid state detectors?

Isn't this device outdone by the "one-shot" HVL measurements using solid state detectors?

Current solid state detectors in the market, to our knowledge, may lack the symmetrical geometric response or construction needed to create the conditions for ideal broad-beam geometry and ideal broad-beam attenuation. Also CT x-ray tubes may have beam filtering materials not properly corrected for or actively compensated for by the solid state detector's construction or software.

How was the idea for the device conceived?

Can I use the standard 100 mm pencil-type CT ionization chamber with the device?

Can I use the standard 100 mm pencil-type CT ionization chamber with the device?

The idea for the MuCT device was conceived in 2009 while conducting research in radiation dose to the skin due to CT-guided procedures. We needed to determine the x-ray beam's effective energy, via the HVL, to perform theoretical skin dose calculations and compare those to experimental radiochromic film results. Unfortunately, due to time

The idea for the MuCT device was conceived in 2009 while conducting research in radiation dose to the skin due to CT-guided procedures. We needed to determine the x-ray beam's effective energy, via the HVL, to perform theoretical skin dose calculations and compare those to experimental radiochromic film results. Unfortunately, due to time constrains, the idea did not materialize. It remained stored inside our brain neurons until 2016 when the idea was approved for our dissertation project. In 2018 the first prototype was fabricated and research data collection commenced to investigate the MuCT device usefulness in determining the HVL in CT scanners without the need to "park" the x-ray tube (i.e. stop the x-ray tube's rotation).

Can I use the standard 100 mm pencil-type CT ionization chamber with the device?

Can I use the standard 100 mm pencil-type CT ionization chamber with the device?

Can I use the standard 100 mm pencil-type CT ionization chamber with the device?

Yes, the MuCT device was designed with the standard 100 mm pencil-type CT ionization chamber in mind. Our research revealed that the chamber's charge collection efficiency drops below 95% along two segments, each one about 10 mm, at the beginning and end of the chamber's active volume, along the chamber's longitudinal axis. This is easily

Yes, the MuCT device was designed with the standard 100 mm pencil-type CT ionization chamber in mind. Our research revealed that the chamber's charge collection efficiency drops below 95% along two segments, each one about 10 mm, at the beginning and end of the chamber's active volume, along the chamber's longitudinal axis. This is easily corrected by repositioning the chamber. Our research also confirmed that using a 300 mm pencil-type chamber completely eliminates any chamber charge collection efficiency issues.

What scanning technique should be used for HVL measurements using the device?

Can I use the standard 100 mm pencil-type CT ionization chamber with the device?

What scanning technique should be used for HVL measurements using the device?

We determined that the optimal scanning technique would be:

  • axial scan
  • slice thickness = 5 mm
  • mAs between 250 and 350
  • one x-ray tube full rotation per second
  • no tube current modulation

Are there any advantages to using the device?

What if I just "park" the x-ray tube and use the traditional narrow-beam geometry ("air-gap") HVL method?

What scanning technique should be used for HVL measurements using the device?

In addition to the capability of measuring the HVL and the x-ray beam's effective energy during clinical scanning conditions, the MuCT device;

  • is light, portable, and easy to use
  • fills the CT beam quality test void in Radiology
  • provides empirical data for computer simulation/models used in radiation dosimetry (ex. Monte Carlo, etc.)
  • enables a

In addition to the capability of measuring the HVL and the x-ray beam's effective energy during clinical scanning conditions, the MuCT device;

  • is light, portable, and easy to use
  • fills the CT beam quality test void in Radiology
  • provides empirical data for computer simulation/models used in radiation dosimetry (ex. Monte Carlo, etc.)
  • enables a more complete/robust equipment performance program
  • can be used to extrapolate the narrow-beam geometry HVL
  • can be used to investigate the off-axis  aluminum equivalent attenuation
  • could allow for potential, yet to be researched, applications in Dual Energy CT, Cone Beam CT, and PET/CT

Current Federal regulations don't require to measure the HVL in CT. Why then should I buy this device?

What if I just "park" the x-ray tube and use the traditional narrow-beam geometry ("air-gap") HVL method?

What if I just "park" the x-ray tube and use the traditional narrow-beam geometry ("air-gap") HVL method?

True, Federal regulations don't require measuring the HVL, in millimeters of aluminum equivalent material, for CT.


We believe that it is not required because the manufacturer's tube inherent filtration specifications, combined with the bow-tie filter attenuation, exceeds Federal x-ray filtration requirements.


We also believe that it is not 

True, Federal regulations don't require measuring the HVL, in millimeters of aluminum equivalent material, for CT.


We believe that it is not required because the manufacturer's tube inherent filtration specifications, combined with the bow-tie filter attenuation, exceeds Federal x-ray filtration requirements.


We also believe that it is not required because the HVL for CT was maybe deemed too difficult to measure, or labor-intensive, because it requires enabling the CT service mode which is not always accessible or easy to do.


Note that the lack of a Federal requirement presents the potential situation where the HVL is never measured during the x-ray tube's or the CT system's service life; an unacceptable situation, in our humble opinion.


Keep in mind that some State regulations could be interpreted as requiring, or clearly mandate, a routine HVL assessment for CT.


The MuCT device is the best solution to any of the issues or situations mentioned above.

What if I just "park" the x-ray tube and use the traditional narrow-beam geometry ("air-gap") HVL method?

What if I just "park" the x-ray tube and use the traditional narrow-beam geometry ("air-gap") HVL method?

What if I just "park" the x-ray tube and use the traditional narrow-beam geometry ("air-gap") HVL method?

We postulate that every point inside the patient, after each x-ray tube's full rotation, could be considered as being irradiated in broad-beam geometry conditions.


The MuCT device, by taking advantage of ideal broad-beam geometry and ideal broad-beam attenuation conditions at the center of rotation (COR), provides an HVL and beam's effecti

We postulate that every point inside the patient, after each x-ray tube's full rotation, could be considered as being irradiated in broad-beam geometry conditions.


The MuCT device, by taking advantage of ideal broad-beam geometry and ideal broad-beam attenuation conditions at the center of rotation (COR), provides an HVL and beam's effective energy assessments that are obtained in realistic clinical conditions.


The traditional narrow-beam geometry ("air-gap") HVL method is time consuming and is not performed in realistic clinical conditions.


Another disadvantage is that the traditional narrow-beam geometry ("air-gap") HVL method may require using the CT service mode which is not always available without the service engineer assistance or requires knowledge of a service access code.

This device would generate lots of scatter in the measurements. How can it allow for an useful HVL calculation?

This device would generate lots of scatter in the measurements. How can it allow for an useful HVL calculation?

This device would generate lots of scatter in the measurements. How can it allow for an useful HVL calculation?

The MuCT device takes advantage of ideal broad-beam geometry and ideal broad-beam attenuation conditions at the COR, and provides an HVL and beam's effective energy assessments that are obtained in clinically relevant and realistic conditions.


Please review additional information available in the "Why it works?" section.

Why not just use a computer model calculation?

This device would generate lots of scatter in the measurements. How can it allow for an useful HVL calculation?

This device would generate lots of scatter in the measurements. How can it allow for an useful HVL calculation?

We can think of several reasons why using the MuCT device is preferable to computer models. Computer models:

  • require expertise in computer programing
  • do not capture individual variations between different units of the same brand and model
  • require proprietary manufacture's data as input, and/or
  • require empirical HVL measurements as input (or f

We can think of several reasons why using the MuCT device is preferable to computer models. Computer models:

  • require expertise in computer programing
  • do not capture individual variations between different units of the same brand and model
  • require proprietary manufacture's data as input, and/or
  • require empirical HVL measurements as input (or for validating the simulated results)


The MuCT device is not affected by any of the issues mentioned above.

Should the MuCT device become a new standard for beam quality in CT?

This device would generate lots of scatter in the measurements. How can it allow for an useful HVL calculation?

Should the MuCT device become a new standard for beam quality in CT?

Only time will tell if Federal agencies, or professional organizations, would mandate, or recommend, the MuCT device as a new standard for beam quality in CT.


However, it is our strong opinion that the MuCT device should indeed become a standard for beam quality in CT.


We envision a routine CT performance program where the MuCT device is us

Only time will tell if Federal agencies, or professional organizations, would mandate, or recommend, the MuCT device as a new standard for beam quality in CT.


However, it is our strong opinion that the MuCT device should indeed become a standard for beam quality in CT.


We envision a routine CT performance program where the MuCT device is used:

  • during the CT acceptance testing, to establish a baseline HVL and effective energy measurements (at the COR), for each kVp and bow-tie filters available
  • to monitor, in a yearly basis, the variation in the baseline HVL and effective energy (at the COR) at least for the most clinically used kVp and bow-tie filter(s)
  • by CT systems manufacturers


Copyright © 2022 MUen LLC - All Rights Reserved.


This website uses cookies.

We use cookies to analyze website traffic and optimize your website experience. By accepting our use of cookies, your data will be aggregated with all other user data.

Accept